The following figure from the SAS-3 specification provides a definition for the Rpre, Rpost, VMA, and
VHL. In each test each of these voltage levels are labeled on the oscilloscope display.
Figure 28Definition of Rpre, Rpost, VMA, and VHL
Rpre is defined as V3/V2.
Rpost is defined as V1/V2.
VMA is defined as V2-V5.
VHL is defined as the peak to peak voltage on the alternating bit portion of the waveforms.
34 922545 Rev A